CONTACT US
DIPTFE Membrane Pleated Filter Cartridge
- Commodity name: DIPTFE Membrane Pleated Filter Cartridge
Description
Product Introduction
DIPTFE series pleated polytetrafluoroethylene filter cartridges use imported hydrophilic modified polytetrafluoroethylene filter membranes as the filtration material. They possess broad chemical compatibility, resisting strong acids and alkalis. The modified hydrophilic properties allow for direct filtration of polar aqueous solutions. 100% integrity testing is performed before shipment to ensure absolute sterilization.

Performance Characteristics
● Excellent high-temperature resistance and broad chemical compatibility.
● Excellent hydrophilicity, suitable for sterilization filtration of strong acid and alkali liquids.
● High flow rate, low pressure drop, and long service life.
● Filter cartridge batch sampling has successfully passed the bacterial challenge test required by regulations, with an LRV >7, and maintains integrity after multiple autoclave sterilizations.
● Each filter cartridge undergoes integrity testing before leaving the factory. All integrity testing limits used have been verified and correlated with bacterial challenge test results.
● All constituent materials comply with the biological testing characteristics of USP Class 6 121°C plastic articles and meet the requirements for food contact materials listed in Chapter 21 of the US Federal Regulations (CFR).
● ISO 9001 and ISO 14001 systems.
● Traceable product production, testing records, and usage.
Filter Cartridge Component Materials
| Filter Media | Hydrophilic Polytetrafluoroethylene (PTFE) Filter Membrane |
| Support/Flow Guide Layer | Polypropylene |
| Mandrel/Outer Cover/End Cap/Point | Polypropylene |
| Sealing Joint/Locking Fins | 222, 226 Polypropylene, with built-in stainless steel reinforcement options |
| Sealing Ring | Silicone Rubber, EPDM Rubber, Fluororubber, etc. |
| Sealing Technology | Adhesive-Free Heat Seal |
Technical Specifications and Usage Conditions
Filter Cartridge Size:
Diameter: 2.75" (69 mm)
Length: 2.5", 5", 10", 20", 30", and 40" sizes, etc.
>Filtration Accuracy:
0.1" μm, 0.22" μm, 0.45" μm
Filtration Area:
Single 10" filter cartridge...
Filter area approximately 0.65 m²
Integrity Test Data Minimum Bubble Point Value:
(Water)
0.22 m ± 0.32 MPa
0.45 m ± 0.24 MPa
Cumulative Steam Sterilization Time:
121℃, 100 hours; 140℃, 40 hours
Hours. Experimental Conditions
Maximum Pressure Difference:
Forward 0.4 MPa, Reverse 0.2 MPa, 25℃
Flow Characteristics

Order Model

Typical Applications
● Sterilization Filtration of Corrosive Liquids
● Sterilization Filtration of Solvents
● Sterilization Filtration of Strong Oxidizing Liquids
● Filtration of Chemical Reagents
Keyword:
DIPTFE Membrane Pleated Filter Cartridge
Get A Quote
Note: Please leave your email address, our professionals will contact you as soon as possible!